Product Details:
|
Warranty: | 18 Months | Test Speed: | Within 15 Minutes |
---|---|---|---|
Storage: | 2℃-30℃ | Sunshine: | Avoid |
Freeze: | Not Allowed | Humidity: | ≤60% |
High Light: | Fast Antigen Influenza AB Test Kit,labnovation Influenza AB Test Kit,20 Test Cassettes fast flu test kit |
Rapid Antigen And Influenza Rapid Test Kit SARS-CoV-2 Antgen And Influenza A/B Rapid Test High Specificity Rapid Test Kit
Intended Use
The rapid test kit is intended for qualitative detection and differentiation of nucleocapsid protein antigens from SARS-CoV-2, influenza type A and influenza type B in the same time directly from nasal/nasopharyngeal (NS/NP) swab, oropharyngeal (OP) swab and saliva specimens from patients with signs and symptoms of respiratory viral infection. The test kit is for professional use only.
Specification
Item |
Antigen test strip performance |
Influenza A test strip performance against PCR | Influenza B test strip performance against PCR |
Sensitivity | 98.03% | 93.3% | 97.00% |
Specificity | 100.00% | 91.00% | 96.40% |
Main Components
Sample Treatment
Add 550μl sample extraction(about 22-24 drops) into the sample tube, dip the swab after collecting the sample into the sample extraction liquid, fully immerse the tip of the swab, rotate and squeeze the swab 10 times, then pull out the swab, and take the stranded liquid as the sample to test cassette.
Use Step
Apply 2 full drops of the treated sample (60μl-70μl) vertically into each of the two sample wells of the test cassette.
Observe the test results immediately within 15~20 minutes, the result is invalid over 20 minutes.
Interpretation Of Result
POSITIVE: The presence of T (T on nCoV /A or B on Flu) and C lines within the reaction window indicate a positive result on SARS-CoV-2 or Flu A and/or B infection or co-infection.
NEGATIVE: One colored line appears in the control region(C). No apparent colored line appears in the test region (T on nCoV /A or B on Flu). The negative result does not indicate the absence of analytes in the sample, it only indicates the level of tested analytes in the sample is less than the minimum detection limit
INVALID: No colored lines appear, or control line fails to appear, indicating that the operator error or reagent failure. Verify the test procedure and repeat the test with a new testing device.
Feature
Cross-reactivity
With human coronavirus 229E, human coronavirus OC43, human coronavirus NL63, adenovirus, human metapneumovirus, parainfluenza virus 1, parainfluenza virus 2, parainfluenza virus 3, parainfluenza virus 4, influenza A, type B Influenza, enterovirus, respiratory syncytial virus, rhinovirus, SARS coronavirus, MERS coronavirus, Haemophilus influenzae, Streptococcus pneumoniae, Streptococcus pyogenes, Candida albicans, Bordetella pertussis, Mycoplasma pneumonia, pneumonia Chlamydia, Legionella pneumophila, etc. have no cross reaction.
With respiratory Syncytial Virus A, measles Virus, Mumps Virus , adenovirus Type 3, parainfluenza Type 2 , partial Pulmonary Virus, human coronavirus OC43, human coronavirus 229E, human herpes virus 6, cytomegalovirus, epstein Barr Virus, enterovirus CA16, rhinovirus, staphylococcus aureus, streptococcus pneumoniae, mycoplasma pneumoniae, parapertussis, chlamydia pneumoniae, escherichia coli, pooled human nasal wash, candida albicans, etc. have no cross reaction.