Product Details:
|
Classification: | IVD Medical Equipment | Method: | HPLC |
---|---|---|---|
Certificated: | NGSP,IFCC | Group: | Universal |
Application: | Daliy Check | Sample Type: | Blood Sample |
HbA1c Analyzer LD-560 HPLC System Fully Automated Analyzer For HbA1c Testing
Intended Use
LD-560 HbA1c analyzer is intended to quantitative determine the HbA1c concentration in the human whole blood and capillary blood.
Specifications
Model | LD-560 |
Method | HPLC |
Parameters | HbA1c, HbF, HbA1a, HbA1b, HbA0; HbA2 (optional) |
Test Speed | 90 seconds per test |
Sample Type | Whole Blood, Pre-diluted blood |
Autoloader Capacity | 20 tubes |
Precision | CV<2% |
Column Capacity | 3000 tests |
Test specimen volume | 5ul (whole blood), 150ul (pre-dilution) |
Feature
Operation Preparation
Before operating this analyzer, please read the operation manual carefully to avoid malfunction or damage.
If the analyzer is used in a manner not specified by the manufacturer, the protection provided by the analyzer may be destroyed.
Before turning on the analyzer, the operator must check the following requirements to ensure that the system is ready.
(1) Check the remaining amount of reagents
Check whether the remaining amount of eluent A, eluent B, eluent C and hemolysis agent is sufficient.
(2) Check the waste liquid barrel
Check whether the waste liquid barrel has been emptied before starting the analyzer every day.
(3) Check the liquid tubes and power supply
Check whether the tubes of reagent and waste liquid are bent and whether the connections are reliable.
Check the analyzer for fluid leakage.
Check whether the power plug of the analyzer is securely inserted into the power socket.
(4) Check the printer
Check whether the paper roll is sufficient and whether it is installed in correct place.