Product Details:
|
Method: | HPLC (High Performance Liquid Chromatography)) | CertificatedWaveLength: | 420nm |
---|---|---|---|
Sample Capacity: | 20 Samples | Parameters: | HbA1c, HbF, HbA1a, HbA1b, HbA0; HbA2 (optional) |
Certificate: | NGSP, IFCC, CE | Working Condition: | Temperature: 15-30℃ Humidity:30%-85% |
Voltage: | 100-240VA | Report Range: | 3-18% |
LD-560 HPLC System for HbA1c Testing
The Fully Automated HbA1c Analyzer is used in conjunction with suitable reagents for the quantitative detection of glycosylated hemoglobin
(HbA1c) in human blood.
Performance Parameteras
Reporting unit: Glycosylated hemoglobin (HbA1c) is reported in NGSP (%) and IFCC units (mmol/mol).
Accuracy: The accuracy is tested with reference materials as samples, and the relative deviation of the
test results should be within ± 5%.
Repeatability: For samples with a test concentration of 4% 6%, the sample coefficient of variation (CV)
of the test results is not greater than 2.0%.
Linear range: Within the range of 3% 18% of glycosylated hemoglobin, the linear correlation coefficien t
R value of the test results should not be less than 0.9900.
Stability: Within 8 hours after the analyzer is turned on and stabilized, the same normal sample is tested, and the relative deviation of the test results is not more than ±3.0%.
Carrying pollution: The carrying pollution rate of the analyzer should not be greater than 2.0%.
Basic function: Provides Chinese report and abnormal alarm.
The temperature accuracy and fluctuation of the chromatography column module: the temperature is within ±1°C of the set value, and the fluctuation is not more than 0.8°C.