Product Details:
|
Principle: | HPLC (High Performance Liquid Chromatography) | Test Speed: | 90s/test |
---|---|---|---|
Sample Types: | Whole Blood Or Pre-diluent Sample | Test Blood Volume: | 5μL In Whole Blood Mode; 150μL In Pre-diluted Mode |
Sample Loading Method: | Continuous Automatic Sample Loading | Sample Capacity: | 20 Samples |
Reagent Pack: | 600T | Column Test Quantity: | 3000T |
Eluent Kind: | Eluent A, Eluent B | Data Storage: | 5000 |
Built-in Thermal Printer: | Yes | CV: | <2% |
Test Range: | 3.0%~18.0% | Operation Screen: | 10" Full Color LCD Display |
Gross Weiht: | 28kg | Dimensions: | 378(L) ×380(W)×510(H)mm |
Shelf Life: | 5 Years | ||
High Light: | Smart Accurate HPLC HbA1c Analyzer,Smart Stable HPLC HbA1c Analyzer,Smart HPLC Glycosylated Hemoglobin Analyzer |
Glycosylated Hemoglobin Diaglosis Fully Automated HbA1c Test HbA1c Analyzer
Intended use
*HPLC, CV<2% - Accurate and Stable
*Full Automatic smart System
*Quality & Performance - IFCC & NGSP Certificate
*User Friendly
Specification
Instrument | Manufacturer | Lanovation |
Manufacture location | Shenzhen, China | |
Model | LD-560 | |
Dection information | Principle method | HPLC |
Wavelength | 420nm | |
Test Speed | 90 Seconds/test | |
Test Speed throughtput | 40 Samples per hour | |
Sample information | Sample types | Whole blood or Pre-diluent sample |
Sample volume | 5μL in whole blood mode 150μL in pre-diluted mode |
|
Sample capacity | 20 samples | |
Sample loading method | Continuous automatic sample loading | |
Mixer | No, but probe mixing samples | |
Reagent and other consumble | Column test quantity | 3000T |
Reagent pack | 600T(Reagent)/ 3000T(Column) separate | |
Eluent | 2 | |
Operation and character information | Operation screen | 10 inch Full color LCD display |
Data storage | 5000 | |
Built-in Thermal Printer | Yes | |
Performance information | CV | <2% |
Test range | 3.0%~18.0% | |
Package Information |
Dimensions | 378(L) ×380(W)×510(H)mm |
Weight | 28kg |
Main Parts
The analyzer consists of an auto sample injection module, a liquid chromatography separation module, a colorimetric detection and control module, a barcode scanning module, and a display printing module.
Reagents and Consumables
Eluent A/Eluent B/Eluent C(Variant Mode) | Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin. |
Hemolysis Reagent | Used to dissolve red blood cells and release hemoglobin. |
Chromatography Column | Used for separation and elution of glycosylated hemoglobin and non glycated. |
Hemoglobin Filter | Used to initially filter the sample after hemolysis. |
Paper Rolls | Prints a report of the test results. |
Certificate
Contact Person: wh@labnovation.com
Tel: +8613480985156