Sample Types: | Whole Blood | Test Blood Volume: | 4μL |
---|---|---|---|
Test Speed Throughtput: | 20 Samples Per Hour | Test Time: | <3 Minutes |
Principle: | Borate Affinity Fluorescent Immunoturbidimetry | Data Storage: | 7000 Tests |
High Light: | Battery Integrated POCT HbA1c Analyzer,Whole Blood POCT HbA1c Analyzer |
LD-100 Analyzer POCT Best Performance IFCC&NGSP Certificated Battery Integrated One Step Use HbA1c Analyzer
Intended Use
Labnovation HbA1c Analyzer LD-100 POCT Analyzer is intended for the percent determination of hemoglobin A1c (HbA1c) in human whole blood which is a semi-automatic instrument used to analyze glycated hemoglobin. LD-100 is small and light, it can be used to detect of HbA1c in medical units at all levels. Fually Automatic system can help the test half work with double results.
Specifications
Model | LD-100 |
Sample Type | Whole blood |
Sample Volume | 4ul |
Test Time | Less than 3 minutes |
Data storage | 7000 |
Method | Fluorescent Immunoturbidimetry |
Operating Type | 1-Step Operation, Fully automatic system |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Voltage | 12VDC-3.33A |
Analytical wave-length | 510nm and 580nm. |
Certification | NGSP, IFCC |
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Performance
Glycated hemoglobin (HbA1c) is reported by default in IFCC units (mmol/mol) and NGSP units (%).
The accuracy is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
Samples with a concentration of 4.0%-6.5% are detected, and the sample coefficient of variation (CV) of the test results is not more than 3%.
In the range of 4%-16% of HbA1c, the linear correlation coefficient r of the test results should be no less than 0.9900.
Within 8 hours after the instrument is turned on and stable, the same normal sample is tested, and POCT HbA1c Analyzer V1.0 2 the relative deviation of the test results is not more than ±3%.
Certificate
Attention
The HbA1c Analyzer meets the emission and immunity requirements specified in the relevant Section of GB/T 18268.26, as shown in the table below.